iVent201 Operator's Manual
213
13. High O
2
% alarm test. Press
Menu and open the Alarm
Settings window. Set the High
O
2
% alarm setting to a value
below the measured value for
FiO
2
(in blue in the alarm
setting bar) and press Accept.
Verify that the High O
2
alarm is
activated on the twentieth
breaths after Accept was
pressed. Close the alarm pop up
window and restore the alarm
setting to its default value.
14. Low O
2
% alarm test. Press
Menu and open the Alarm
Settings window. Set the Low
O
2
% alarm setting to a value
above the measured value for
FiO
2
(in blue in the Alarm
setting bar) and press Accept.
Verify that the Low O
2
% alarm
is activated on the tenth breath
after Accept was pressed. Close
the alarm pop up window and
restore the alarm setting to its
default value.