XT2640OperatingManual 13July2016 Page155of187
SubField
Subfield
Format
Value Description
STBYERR Respondswithstandbypowermeasurementerror(in
W
)
MeasurementSource KEYWORD
CH1,CH2,CH3or
CH4
CH1isDEFAULTifnootherMeasurementSourcesub‐fieldfound
Dataisfromtheselectedchannel
A1,A2,A3,VPA1,
VPA2orVPA3
DataisfromtheselectedVPA
MOTOR
Dataisamotormeasurementresult(onlytheDatafieldisusedwith
thissource,anyotherfieldsareignored)
IN
DataisfromtheINefficiencygroup(orstartsatthisgroupif2
nd
MeasurementSourceisMIDDLEorOUT)
MIDDLE
DataisfromtheMIDDLEefficiencygroup(orstartsatthisgroupif
2
nd
MeasurementSourceisOUT)
OUT DataisfromtheOUTefficiencygroup
2
nd
MeasurementSource(ignored
ifMeasurementSourceisasingle
channelorisMOTOR)
KEYWORD
MIDDLE
DataistotheMIDDLEefficiencygroup(onlyvalidifMeasurement
SourceisIN)
OUT
DataistotheOUTefficiencygroup(onlyvalidifMeasurement
SourceisINorMIDDLE)
pA…D Dataisfromthe1
st
through4
th
(resp.)channeloftheselectedVPA
pAC
DataisthevoltagemeasuredbetweenphasesAandCoftheselected
VPA
pAB
DataisthevoltagemeasuredbetweenphasesAandBoftheselected
VPA
pBC
DataisthevoltagemeasuredbetweenphasesBandCoftheselected
VPA
pN DataistheneutralcurrentoftheselectedVPA
WYE Dataisthe‘Wye’voltageoftheselectedVPA
DELTA Dataisthe‘Delta’voltageoftheselectedVPA
SEQZERO DataisthezerosequencedataoftheselectedVPA
SEQPOS DataisthepositivesequencedataoftheselectedVPA
SEQNEG DataisthenegativesequencedataoftheselectedVPA
TOTALor
AVERAGE
DEFAULTifnoother2
nd
MeasurementSourcesub‐fieldfound
DataisthetotalfortheselectedVPA,exceptforVOLTSandAMPS
datawhichistheaverageofallchannels/phases
MeasurementType KEYWORD
DC DataistheDCcomponent
AC DataistheACcomponen
t
ACDC orRMS DataistheACDCcomponen
t
COUPLED
DEFAULTifnootherMeasurementTypesub‐fieldfound
DataistheDC,ACorACDCcomponentasCOUPLINGsetting
RECTIFIED Dataistheaveragerectified (ACDC)
STBY Dataistheaveragestandbypowermeasuremen
t
(ACDC)
STBYMIN Dataistheminimumstandbypowermeasuremen
t
(ACDC)
STBYMAX Dataisthemaximumstandbypowermeasuremen
t
(ACDC)
STBYCFMIN
Dataistheminimumcrestfactorstandbypowermeasuremen
t
(ACDC)
STBYCFMAX
Dataisthemaximumcrestfactorstandbypowermeasurement
(ACDC)
STBYTHDMAX
DataisthemaximumTHDstandbypowermeasurement(asthe
percentageofthefundamental)
STBYSLOPE Dataistheslope(inW/sec)ofthestandbypowermeasurement
FF Dataistheformfactor(ACDC)
CF Dataisthecrestfactor(ACDC)
PK Dataispeak(ACDC)
VALLEY Dataisvalleypeak(ACDC)
PK‐VLY Dataisthedifferencebetweenpeakandvalleypeak
HIPK Dataisthehighestpeak(includingpolarity,ACDC)
LOPK Dataisthelowestpeak(includingpolarity,ACDC)
THDF DataisTHDrelativetofundamentalamplitude(in%)
THDSIG DataisTHDrelativetoACDCamplitude(in%)
H1…500 Dataisharmonicamplitudedataforthespecifiedharmonic
P1…500
Dataisharmonicphasedataforthespecifiedharmonic(indegrees
with±180°range)
INRUSH Dataistheinrushdata(maxhold,ACDC)
INRUSHPK Dataisthepeakinrushdata(maxhold,PKdata)
INRUSHVLY Dataisthe valley peakinrushdata(minhold,VALLEYdata)
INTEG Dataistheintegrateddata (asCOUPLINGsetting)
INTAVG Dataistheaverageintegrateddata(asCOUPLINGsetting)
CHARGE Dataistheintegratedchargedata (DConly)
DISCHARGE Dataistheintegrateddischargedata(DConly)
BOUGHT Dataistheintegratedboughtdata (asCOUPLINGsetting)
SOLD Dataistheintegratedsolddata (asCOUPLINGsetting)
%2…500 DataisasHnbutexpressedasapercentageofthefundamental
%S2…500 DataisasHnbutexpressedasapercentageoftheACDCamplitude
THC
TotalHarmonicCurrent(onlyvalidwhenusedwithAMPSdatafora
channel)

Table of Contents