XT2640OperatingManual 13July2016 Page155of187
Sub‐Field
Sub‐field
Format
Value Description
STBYERR Respondswithstandbypowermeasurementerror(in
)
MeasurementSource KEYWORD
CH1,CH2,CH3or
CH4
CH1isDEFAULTifnootherMeasurementSourcesub‐fieldfound
Dataisfromtheselectedchannel
A1,A2,A3,VPA1,
VPA2orVPA3
DataisfromtheselectedVPA
MOTOR
Dataisamotormeasurementresult(onlytheDatafieldisusedwith
thissource,anyotherfieldsareignored)
IN
DataisfromtheINefficiencygroup(orstartsatthisgroupif2
nd
MeasurementSourceisMIDDLEorOUT)
MIDDLE
DataisfromtheMIDDLEefficiencygroup(orstartsatthisgroupif
2
nd
MeasurementSourceisOUT)
OUT DataisfromtheOUTefficiencygroup
2
nd
MeasurementSource(ignored
ifMeasurementSourceisasingle
channelorisMOTOR)
KEYWORD
MIDDLE
DataistotheMIDDLEefficiencygroup(onlyvalidifMeasurement
SourceisIN)
OUT
DataistotheOUTefficiencygroup(onlyvalidifMeasurement
SourceisINorMIDDLE)
pA…D Dataisfromthe1
st
through4
th
(resp.)channeloftheselectedVPA
pAC
DataisthevoltagemeasuredbetweenphasesAandCoftheselected
VPA
pAB
DataisthevoltagemeasuredbetweenphasesAandBoftheselected
VPA
pBC
DataisthevoltagemeasuredbetweenphasesBandCoftheselected
VPA
pN DataistheneutralcurrentoftheselectedVPA
WYE Dataisthe‘Wye’voltageoftheselectedVPA
DELTA Dataisthe‘Delta’voltageoftheselectedVPA
SEQZERO DataisthezerosequencedataoftheselectedVPA
SEQPOS DataisthepositivesequencedataoftheselectedVPA
SEQNEG DataisthenegativesequencedataoftheselectedVPA
TOTALor
AVERAGE
DEFAULTifnoother2
nd
MeasurementSourcesub‐fieldfound
DataisthetotalfortheselectedVPA,exceptforVOLTSandAMPS
datawhichistheaverageofallchannels/phases
MeasurementType KEYWORD
DC DataistheDCcomponent
AC DataistheACcomponen
ACDC orRMS DataistheACDCcomponen
COUPLED
DEFAULTifnootherMeasurementTypesub‐fieldfound
DataistheDC,ACorACDCcomponentasCOUPLINGsetting
RECTIFIED Dataistheaveragerectified (ACDC)
STBY Dataistheaveragestandbypowermeasuremen
(ACDC)
STBYMIN Dataistheminimumstandbypowermeasuremen
(ACDC)
STBYMAX Dataisthemaximumstandbypowermeasuremen
(ACDC)
STBYCFMIN
Dataistheminimumcrestfactorstandbypowermeasuremen
(ACDC)
STBYCFMAX
Dataisthemaximumcrestfactorstandbypowermeasurement
(ACDC)
STBYTHDMAX
DataisthemaximumTHDstandbypowermeasurement(asthe
percentageofthefundamental)
STBYSLOPE Dataistheslope(inW/sec)ofthestandbypowermeasurement
FF Dataistheformfactor(ACDC)
CF Dataisthecrestfactor(ACDC)
PK Dataispeak(ACDC)
VALLEY Dataisvalleypeak(ACDC)
PK‐VLY Dataisthedifferencebetweenpeakandvalleypeak
HIPK Dataisthehighestpeak(includingpolarity,ACDC)
LOPK Dataisthelowestpeak(includingpolarity,ACDC)
THDF DataisTHDrelativetofundamentalamplitude(in%)
THDSIG DataisTHDrelativetoACDCamplitude(in%)
H1…500 Dataisharmonicamplitudedataforthespecifiedharmonic
P1…500
Dataisharmonicphasedataforthespecifiedharmonic(indegrees
with±180°range)
INRUSH Dataistheinrushdata(maxhold,ACDC)
INRUSHPK Dataisthepeakinrushdata(maxhold,PKdata)
INRUSHVLY Dataisthe valley peakinrushdata(minhold,VALLEYdata)
INTEG Dataistheintegrateddata (asCOUPLINGsetting)
INTAVG Dataistheaverageintegrateddata(asCOUPLINGsetting)
CHARGE Dataistheintegratedchargedata (DConly)
DISCHARGE Dataistheintegrateddischargedata(DConly)
BOUGHT Dataistheintegratedboughtdata (asCOUPLINGsetting)
SOLD Dataistheintegratedsolddata (asCOUPLINGsetting)
%2…500 DataisasHnbutexpressedasapercentageofthefundamental
%S2…500 DataisasHnbutexpressedasapercentageoftheACDCamplitude
THC
TotalHarmonicCurrent(onlyvalidwhenusedwithAMPSdatafora
channel)