17-4 IM 765501-01E
Errors in Execution (-200 to -299)
Error No. Error Message Corrective Action Page
–221 Setting conflict The following four setting conflicts are possible. -
• Pulse source and block average
• Pulse source and auto measurement range
• The source and measurement are set to the same function
and using auto measurement range
• Limiter tracking OFF and auto measurement range
Change the settings so that these conflicts do not occur.
–222 Data out of range Check the selectable range of the parameter. -
If the command can use MINimum and MAXimum as its
parameter, the range can also be queried.
–256 Filename not found Check that the file exists. 16-16, 16-27,
You can also use the CATalog? command to query the list of 16-34
files.
–285 Program syntax error Check that the sweep pattern file is in the correct format. 6-11
Device Errors (-300 to -399)
Error No. Error Message Corrective Action Page
–350 Queue overflow Read the error using :SYSTem:ERRor? or clear the error 16-35, 16-40
queue using *CLS.
–361 Parity error Check that the communication settings on the GS610 and 14-6
PC match.
If the settings are correct, check the cable, and lower the baud rate.
–362 Framing error Check that the communication settings on the GS610 and 14-6
PC match.
If the settings are correct, check the cable, and lower the baud rate.
–363 Input buffer overrun Set the handshaking to a setting other than OFF. 14-6
Lower the baud rate.
Query Errors (-400 to -499)
Error No. Error Message Corrective Action Page
–410 Query INTERRUPTED Check transmission/reception procedure. 16-3
–420 Query UNTERMINATED Check transmission/reception procedure. 16-3
–430 Query DEADLOCK Keep the program message shorter than or equal to 64 KB. 16-3
17.2 Error Code Description and Corrective Action