7-20
IM 701310-01E
When Nogo Is Set as an Action and Interval Is Not Set to OFF
<Subject>: The subject of the e-mail. The content of the subject is as
follows:
GoNogo Interval Report (No.).
The number in parentheses is the number of returned no-go
results.
[Comment]: Comments
[Setup Information]: Information about GO/NO-GO conditions 1 to 4.
Logic (AND/OR)
Stop Nogo/ACQ Count (The number of no-go results/the
acquisition count)
[TimeRange]: The amount of time that has elapsed since the start of data
acquisition
[Nogo/ExecCount]: The number of no-go results/the number of determinations
[EachNogoCount]: The number of no-go results for each GO/NO-GO condition
(when using OR logic).
This is not displayed when
AND logic is used.
Example of a Typical E-Mail
Subject GoNogo Triggered Report 23
--------------------------------------------------Body
[Comment] Sample -GoNogo
[Setup Information] Sel
ect1:Rect(C1,Main) (Left:-3.0000E+00,Right:-2.5000E+00,Upper: 500.00E-03,Lower:-500.00E-03,Condition:In)
Select2:Wave(C2,Z1)
(Range1:-5.0000E+00,Range2: 5.0000E+00,Condition:Out)
Select3:Polygon(C3,Z2)
(Condition:In)
Select4:Measure(Max(C4))
(Upper: 1.0000E+00,Lower:-1.0000E+00,Condition:Out)
Logic:OR
Stop Nogo/ACQ Count:100/100
[Trigger Date and Time] 2006/03/06 13:53:46
[Nogo/Exec Count] 23/56
[Nogo Factor] Select1:Rect(C1,Main)
Select2:Wave(C2,Z1)
Select3:Polygon(C3,Z2)