Acquisition and Display
7-41
IM 701310-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
App
Index
Setting the Determination Mode and Category
5.
Press these soft keys: Mode > Parameter.
This sets the determination mode to Parameter.
6.
Press these soft keys: Category > FFT.
This sets the parameter category to FFT.
Selecting the Analysis Item/Calculation
Selecting the Source Window
7.
Press the Window soft key to select FFT1 (the results of the analysis of
Window1) or FFT2 (the results of the analysis of Window2).
Selecting the Analysis Item
8.
Press the Item soft key.
9.
Press the Basic or Calc soft key.
If you select Basic, proceed to step 10.
If you select Calc, proceed to step 11.
10.
Use the rotary knob and SET to select the waveform parameters.
Proceed to step 12.
7.14 Setting FFT Parameter GO/NO-GO Determination Conditions