Setting the Determination Mode and Category
5.
Press these soft keys: Mode > Parameter.
This sets the determination mode to Parameter.
6.
Press these soft keys: Category > XY.
This sets the parameter category to X-Y waveform parameters.
Selecting the Source Window
7.
Press the Window soft key to select XY1 (the X-Y waveform of Window1) or XY2
(the X-Y waveform of Window2).
Setting the GO/NO-GO Determination Range
8.
Press the Upper/Lower soft key to select the direction of the range that you will
set.
9.
Use the rotary knob to set the range’s maximum and minimum values.
7.15 Setting X-Y Waveform Parameter GO/NO-GO Determination Conditions