10-73
IM 701310-01E
Analysis and Search
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
App
Index
Level
Set the reference level for CH1 to CH4 and M1 to M4.
The selectable range is �10 divisions from the vertical position. The resolution is 0.01
divisions. For example, if the T/div setting is 2 mV/division, the resolution is 0.02 mV.
Hysteresis
Set the hysteresis for CH1 to CH4 and M1 to M4.
The selectable range is from 0.0 to 4.0 divisions. The resolution is 0.1 divisions.
Trigger hysteresis settings
and correspond to 0.6 divisions and 1.0 division.
Found Point
The found-point position is the same as the trigger point position. For a description of the
trigger point, see the example in “Explanation” in section 6.14.
10.12 Searching Serial Pattern Signals