4-14
IM MX100-01E
Maximum input voltage:
DC voltage at 1-V range or less, TC, RTC, and DI (contact): VDC
(continuous)
Othermeasurementranges:±120VDC(continuous)
Normal-mode voltage:
DC voltage, TC, DI (LEVEL): 1.2 times the range rating or less
(50/60 Hz, peak value including the signal component)
RTD100Ω:50mVpeak
RTD10Ω,25Ω,50Ω:10mVpeak
Normal-mode rejection ratio:
For integration time of 16.67 ms or more: 40 dB or more (50/60
Hz±0.1%)
For integration time of 1.67 ms: 50/60 Hz is not rejected
The RTD ranges indicate the voltage conversion value when
measurement current applied.
Common-mode voltage:
600 VACrms (50/60 Hz), reinforced (double) insulation
Common-moderejectionratio(50/60Hz±0.1%,500Ωunbalanced,betweentheminus
terminal and ground)
For integration time is 16.67 ms or more: 120 dB or more
For integration time is 1.67 ms or more: 80 dB or more
The RTD ranges indicate the voltage conversion value when
measurement current applied.
Common-mode voltage between channels:
120 VACrms (50/60 Hz)
Noise rejection: Rejection by the integrating A/D converter and the use of low pass
lters
Inputresistance: DCvoltageat1Vrangeorless,TCrange:10MΩormore
DCvoltageat2Vrangeormore:Approx.1MΩ
Insulationresistance: 20MΩormore(500VDC)betweentheinputandground
Input bias current: 10 nA or less (except for burnout setting)
Withstand voltage: 1000 VAC (50/60 Hz) for one minute between input terminals
3700 VAC (50/60 Hz) for one minute between the input terminal
and earth terminal
Input signal source resistance:
DCvoltageandTC:2kΩorless
RTD50Ωand100Ωsystems:10Ωorlesspercable
RTD10Ωand25Ωsystems:1Ωorlesspercable
Thermocouple burnout:
Detectionataspecieddetectionintervalpermeasurement
interval and detection within the thermocouple range (detection
ON/OFF possible)
Up/Down setting is possible
2kΩorless:normal,200kΩormore:linebroken
(Shunt capacity of 0.01 µF or less), detection current of
approximately 10 µA, and detection time of approximately 2 ms
Shunt capacity during RTD: 0.01 mF or less
Power consumption: Approximately 1.2 W
External dimensions: Approximately 57 (W) × 131 (H) × 151 (D) mm (including the
terminal cover)
Weight: Approximately 0.5 kg
Terminal type: Clamp, terminal board is detachable
Applicable wire size: 0.14 to 1.5 mm
2
(AWG26 to 16)
4.5 10-CH, Medium-Speed Universal Input Module Specifications