Chapter 11 Monitoring Invasive Pressures (IBP)
11-2 www.zoll.com 9650-001355-01 Rev. M
Warning! If electrocautery is used, always avoid using any transducer with a conductive
(metal) case connected to its ground shield. Using a conductive transducer case that
is connected to its cable shield risks high-frequency burns at the ECG electrodes if
the transducer case becomes grounded to earth.
Normal alarm functions will detect complete disconnections of invasive pressure
transducers; however, the alarm functions will not detect a partial disconnection or
the use of some incompatible transducers. Use only approved transducers and ensure
that the transducers are connected properly.
Before you use the X Series unit on a new patient, always turn it off for at least 2
minutes. This clears the previous patient’s trend values, alarm limit settings, and
NIBP cuff inflation pressure.
•Use only ZOLL-approved IBP sensors. Use of non-approved sensors may result in
inaccurate IBP measurements.
IBP Setup
To take safe and accurate IBP measurements using the X Series, you must perform the
following steps, each of which corresponds to a section in this chapter. Read each section
carefully before you perform IBP measurements.
1. Attach the invasive pressure transducer to the X Series unit.
2. Zero the transducer.
3. Set the invasive pressure alarms (according to your organization’s standards).
4. Select a label for the invasive pressure channel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL X Series and is the answer not in the manual?

ZOLL X Series Specifications

General IconGeneral
TypeDefibrillator/Monitor
DisplayColor LCD
ECG MonitoringYes
PacingYes
SpO2 MonitoringYes
NIBP MonitoringYes
Defibrillator TypeBiphasic
EtCO2 MonitoringOptional
Data StorageYes
BatteryRechargeable Lithium-ion
CPR FeedbackReal CPR Help
Energy Selection1-360 J
Pacing Rate30 – 180 ppm
Pacing Output0 – 140 mA
Display Size6.5 inches
Invasive Pressure MonitoringOptional
Temperature MonitoringOptional
Wireless ConnectivityWi-Fi, Bluetooth

Related product manuals