X Series Product Functions
9650-001355-01 Rev. M X Series Operator’s Guide 1-13
Batteries
X Series models use an easily replaced rechargeable lithium-ion battery pack (the SurePower II
Battery Pack). A new, fully charged battery pack typically delivers more than 6 hours of ECG
monitoring. Use of other functions (such as the defibrillator, printer, or pacemaker) reduces this
time.
When a LOW BATTERY icon appears on the display and the unit emits three beeps in
conjunction with the displayed battery icon, the battery must be replaced and recharged.
You can charge the battery by either of the following methods:
Internal charging — plug the X Series into an auxiliary power adapter to automatically
begin charging the installed battery pack. The front panel battery indicator operates as
follows:
Note: Upon power up, it takes approximately 45 seconds for the LEDs on the battery to
accurately display run time.
External charginguse the ZOLL SurePower Battery Charger with the X Series battery
adapter to charge the battery pack and test the battery’s capacity. For details, refer to the
SurePower II Battery Pack Guide.
The Recalibration LED icon ( ) lights for approximately 10 seconds (after you press and
release the Display button) if the battery needs to be calibrated. If the Recalibration LED lights,
the runtime indicator will not display run time for that battery. For best performance of the
battery, you should recalibrate the battery as soon as possible.
To manually recalibrate the SurePower II Battery Pack, you can insert the battery into the
SurePower Charger Station and perform a Manual Test (for more information, see the ZOLL
SurePower Charger Station Operators Guide).
After you recalibrate the battery, the Recalibration LED will only flash when you press the
Display button.
When the indicator is: It means:
Steady yellow Battery is charging.
Steady green Battery is charged.
Alternating yellow and
green
The charge state cannot be
determined or a battery charging
fault has been detected.
Not lit No battery in device.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL X Series and is the answer not in the manual?

ZOLL X Series Specifications

General IconGeneral
TypeDefibrillator/Monitor
DisplayColor LCD
ECG MonitoringYes
PacingYes
SpO2 MonitoringYes
NIBP MonitoringYes
Defibrillator TypeBiphasic
EtCO2 MonitoringOptional
Data StorageYes
BatteryRechargeable Lithium-ion
CPR FeedbackReal CPR Help
Energy Selection1-360 J
Pacing Rate30 – 180 ppm
Pacing Output0 – 140 mA
Display Size6.5 inches
Invasive Pressure MonitoringOptional
Temperature MonitoringOptional
Wireless ConnectivityWi-Fi, Bluetooth

Related product manuals