Chapter 12 Monitoring Temperature
12-2 www.zoll.com 9650-001355-01 Rev. M
Selecting and Applying Temperature Probes
You should use only temperature probes that are approved for use with the X Series unit. See
Appendix B, Accessories for a list of ZOLL-approved temperature probes. The use of other
probes that do not match the performance specifications of the ZOLL-approved probes may
produce incorrect temperature readings.
To apply the temperature probe to the patient, follow your organization’s standard procedures.
Always refer to probe manufacturers Directions for Use before using the probe.
Warning! The application and use of metal-jacketed temperature probes that come in contact
with conductive objects or clinical personnel during electrocautery may cause burns
at the point of contact between the patient and the temperature probe.
To ensure safe and reliable operation including biocompatibility, use only ZOLL-
approved temperature probes.
Connecting the Temperature Probe
To connect the temperature probe, insert the probe’s 1/4” plug into one of the two connection
jacks on the side of the X Series unit.
Figure 12-1 Connecting the Temperature Probe to the X Series Unit
Displaying Temperature
When you connect the probe, the unit displays the temperature after a brief pause. The X Series
unit displays temperature as a numeric value in the Temperature window. You can specify that
the unit display the temperature in
C or F.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL X Series and is the answer not in the manual?

ZOLL X Series Specifications

General IconGeneral
TypeDefibrillator/Monitor
DisplayColor LCD
ECG MonitoringYes
PacingYes
SpO2 MonitoringYes
NIBP MonitoringYes
Defibrillator TypeBiphasic
EtCO2 MonitoringOptional
Data StorageYes
BatteryRechargeable Lithium-ion
CPR FeedbackReal CPR Help
Energy Selection1-360 J
Pacing Rate30 – 180 ppm
Pacing Output0 – 140 mA
Display Size6.5 inches
Invasive Pressure MonitoringOptional
Temperature MonitoringOptional
Wireless ConnectivityWi-Fi, Bluetooth