Chapter 12 Monitoring Temperature
12-6 www.zoll.com 9650-001355-01 Rev. M
Temperature System Messages
The X Series unit may display the following messages when monitoring Temperature.
Note: The temperature function performs a self test when initially powered on and also
performs system tests automatically, every 10 seconds, while this function is active.
System Message Cause
CHECK PROBE The temperature probe is disconnected. Check probe
and reconnect it.
PROBE FAILURE The temperature probe is defective. Replace the
Temperature probe.
TEMP DISABLED A system error has occurred. The X Series unit cannot
take temperature measurements and should be serviced.