Chapter 12 Monitoring Temperature
12-4 www.zoll.com 9650-001355-01 Rev. M
Setting Upper and Lower Temperature Alarm Limits
Initially, the Temperature Alarm Settings menu specifies that the Temperature alarms are
enabled (ON), and displays the default upper and lower limits. The upper and lower limits can
be ON or OFF (default is OFF). The following table lists the default Temperature limits for
adult, pediatric, and neonate patients, and gives the range in which you can set these limits:
Patient Type
Temperature Limit Default Temperature Limit Range
Adult Lower: 0.0F (0.0° C)
Upper: 5.0F (-15° C)
Lower: 0.0 - 89.8 F (0.0 - 32.1° C)
Upper: 0.2 - 90.0 F (-17.7 - 32.2° C)
Pediatric Lower: 0.0F (0.0° C)
Upper: 5.0F (-15° C)
Lower: 0.0 - 89.8 F (0.0 - 32.1° C)
Upper: 0.2 - 90.0 F (-17.7 - 32.2° C)
Neonate Lower: 0.0F (0.0° C)
Upper: 5.0F (-15° C)
Lower: 0.0 - 89.8 F (0.0 - 32.1° C)
Upper: 0.2 - 90.0 F (-17.7 - 32.2° C)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL X Series and is the answer not in the manual?

ZOLL X Series Specifications

General IconGeneral
TypeDefibrillator/Monitor
DisplayColor LCD
ECG MonitoringYes
PacingYes
SpO2 MonitoringYes
NIBP MonitoringYes
Defibrillator TypeBiphasic
EtCO2 MonitoringOptional
Data StorageYes
BatteryRechargeable Lithium-ion
CPR FeedbackReal CPR Help
Energy Selection1-360 J
Pacing Rate30 – 180 ppm
Pacing Output0 – 140 mA
Display Size6.5 inches
Invasive Pressure MonitoringOptional
Temperature MonitoringOptional
Wireless ConnectivityWi-Fi, Bluetooth

Related product manuals