Heart Rate Meter
9650-001355-01 Rev. M X Series Operator’s Guide 7-5
Heart Rate Meter
The Heart Rate meter displays the heart rate that it derives from the ECG monitoring function
(by default) or from a monitoring function that you specify. If the ECG (or user-specified
monitor function) measurements are not available, the Heart Rate meter derives the heart rate
from the following monitoring functions, if they are available, in this order: User-selected
default source, ECG, IBP channel 1, SpO
2
, IBP channel 2, IBP channel 3, and NIBP. The Heart
Rate meter is labeled
HR (as in the following example) if the source is ECG, and PR if any
other source is used.
Note: The Heart Rate meter displays heart rates below 20 as three dashes (
---) and heart rates
above 300 as three plus signs (
+++).
Configuring Heart Rate (HR) Meter Alarms
The X Series unit allows you to enable and disable the Heart Rate (HR) alarm, to set alarm
limits, and to select a Heart Rate tone.
Enabling/Disabling HR Alarms and Setting Alarm Limits
When enabled, the X Series unit sounds alarms whenever the patient’s heart rate is above or
below the specified heart rate alarm limits.
To enable (or disable) HR alarms and set Upper and Lower alarm limits, you can either do so
through the Alarms quick access key ( ) or the HR/PR Parameter Control Panel.
To configure HR alarms through the Alarms quick access key:
1. Press .
2. Press .
3. Press .
4. Use the navigation buttons to highlight and select HR/PR Alarm.
5. On the HR/PR Alarm Settings menu, use the navigation keys to select the fields that you want
to change:
• Upper Limit Enable
• Lower Limit Enable
• Upper Limit
• Lower Limit
When you are finished changing values on the alarm menu, navigate to the Backarrow key to
confirm your choices and exit the menu.