9650-001355-01 Rev. M X Series Operator’s Guide 24-1
Chapter 24
Maintenance
Resuscitation equipment must be maintained to be ready for immediate use. To ensure the
readiness and optimum working condition of the X Series unit, you should perform the
following inspections and tests daily or at each shift change.
In addition to the daily check, authorized personnel should complete performance and
calibration testing at regularly scheduled intervals, which should not exceed one year.
A maintenance log is an important part of a successful maintenance program in which you
record information on a regular basis. This allows for verification of necessary maintenance
and for scheduling periodic requirements such as calibration and certification.
In accordance with the recommendations of the Defibrillator Working Group of the Food and
Drug Administration, an operator’s shift check list is included in this chapter, which you can
copy for use as needed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL X Series and is the answer not in the manual?

ZOLL X Series Specifications

General IconGeneral
TypeDefibrillator/Monitor
DisplayColor LCD
ECG MonitoringYes
PacingYes
SpO2 MonitoringYes
NIBP MonitoringYes
Defibrillator TypeBiphasic
EtCO2 MonitoringOptional
Data StorageYes
BatteryRechargeable Lithium-ion
CPR FeedbackReal CPR Help
Energy Selection1-360 J
Pacing Rate30 – 180 ppm
Pacing Output0 – 140 mA
Display Size6.5 inches
Invasive Pressure MonitoringOptional
Temperature MonitoringOptional
Wireless ConnectivityWi-Fi, Bluetooth