Alarm Options
9650-001355-01 Rev. M X Series Operator’s Guide 5-9
SpO
2
(%)
Entire range Limit = 100 (Adult and
Pediatric)
Limit = Numeric + 5 (Neo-
nate)
Limit = Numeric – 5
SpCO
(%)
Entire range Limit = Numeric + 2
Upper limit 40
Limit = Numeric - 2
Lower limit 0
SpMet
(%)
Entire range Limit = Numeric + 2
Upper limit 15
Limit = Numeric - 2
Lower limit 0
SpHb
(g/dL)
Entire range Limit = Numeric + 2 Limit = 0
SpHb
(mmol/L)
Entire range Limit = Numeric + 1 Limit = 0
SpOC
(ml/dL)
Entire range Limit = Numeric + 2 Limit = 0
PVI
(%)
Entire range Limit = Numeric + 5 Limit = Numeric - 5
PI
(%)
Entire range Limit = Numeric + 2 Limit = 0
EtCO
2
(mmHg)
Entire range Limit = Numeric + 10 Limit = Numeric - 5 mmHg
unless the numeric falls
below the lower alarm limit
range, in which case Stat
Set sets the lower limit to
15 mmHg.
FiCO
2
(mmHg)
Entire range Limit = Numeric + 5 N/A
Temp
(°C)
Entire range Limit = Numeric + 0.5 Limit = Numeric – 0.5
Temp
(°F)
Entire range Limit = Numeric + 0.9 Limit = Numeric – 0.9
Parameter
(units)
Range Upper Limit Calculation Lower Limit Calculation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL X Series and is the answer not in the manual?

ZOLL X Series Specifications

General IconGeneral
TypeDefibrillator/Monitor
DisplayColor LCD
ECG MonitoringYes
PacingYes
SpO2 MonitoringYes
NIBP MonitoringYes
Defibrillator TypeBiphasic
EtCO2 MonitoringOptional
Data StorageYes
BatteryRechargeable Lithium-ion
CPR FeedbackReal CPR Help
Energy Selection1-360 J
Pacing Rate30 – 180 ppm
Pacing Output0 – 140 mA
Display Size6.5 inches
Invasive Pressure MonitoringOptional
Temperature MonitoringOptional
Wireless ConnectivityWi-Fi, Bluetooth