EasyManua.ls Logo

ABB Freelance 2000 - C 4.3.3 Area Options Tab

ABB Freelance 2000
226 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section C Standard Displays
C-19
C 4.3.3 Area options tab
Trend area options The parameter definition of position comprises the time range for depiction
on the X-axis, the definition of scaling on the Y-axis and band start / band
end for up to six trends.
Time gaps Maximum time interval between two successive time stamps of a value.
The values in this internal are interpolated (see Operators Manual, Ope-
rator Station, Trend display, Interpolation). Entry is made acc. to IEC
1131-3 time format.
Time axis The input field must be filled in.
The time range specifies the yardstick on the X-axis for the trend depic-
tion in DigiVis. Entry is made acc. to IEC1131-3 time format. The highest
value amounts to about T#24d20h31m23s.
Example:
T#2147s or T#24d20h31m23s
Band start The Band start specifies the value for scale start of the Y-axis for depict-
ing an individual trend. See figure below.
Values entered as real numbers.
The value range is 0.0 and ±9999999999.9. The value for the band start
must be less than the value for the band end.
% The percentage value specifies the position of the scale start for Band
start % on the Y-axis. See figure below.
The percentage values must be entered as integers. The value range is
0 - ±5000%.
tt0432us.bmp
Gross Automation, 1725 South Johnson Road, New Berlin, WI 53146, www.ssacsales.com, 800-349-5827

Table of Contents