Adixen ASM GRAPH - Procedure with User Level; Adjust Vacuum;Sniffing Alarm Reject Point; Purpose of the Sniffer Probe Clogged Reject Point

Default Icon
336 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C
211
Operation of the leak detector
3/4
GB 01362 -
Edition
02 -
September
04
Purpose
of the sniffer probe
clogged reject point
When the helium signal is lower than the
set «sniffer
probe clogged»
reject point, an information will be actived.
Adjustment
2x
C
110
Adjust
vacuum/snif
fing
alarm reject point
Note:
Adjustment process
is different depending on
user
interface level
of the operator:
For operator with
user
interface level
1
or
2
:
access
to level
4
with password.
To access
to level
4
C
130
Procedure
with user level
3
Hard vac.
test
Adjustment
C
110
Sniffing test
Procedure
with user level
4
Adjustment
Hard vac. test
C
110
Adjustment
Sniffing test
The
unit
selection
of the leak
detector
is the
threshold
unit. If the operator
would like to
change
unit, he
must convert
the
threshold
unit and
reprogrammed it with the new one.