C
303
External calibration of the leak detector
3/4
GB 00262
-
Edition 05
-
September
06
Sniffing mode
Hard
vacuum
mode
T
arget
value
(*)
adjustment
manually
automatically
Adjustment
C C
304
COR
(*)
The
target value is the desired value to be displayed on the
digital
display at the end of the calibration..
The
target value can be memorized following 2 methods:
the operator
enters
directly the target value,
the target value is automatically calculated by the leak detector.
About sniffing
test
mode
In sniffing
test
mode, the calibration can be performed with:
the ambient atmosphere,
an external calibrated leak,
a container or installation, at
atmospheric
pressure,
filled with a known gas mixture including Helium.
The
sniffer probe will be exposed to one of the 3 items
listed above (3rd one
represents
the
most
reliable and accurate
way
of calibrating a leak
detector
in sniffing mode).
The
l
st
and 2nd
ones
are the
most common
and practical methods.
Before
using one of the 2
items, connect
the sniffer probe to the sniffer
port of the leak detector.