External calibration of the leak detector
l/4
GB 00262
-
Edition 05
-
September
06
A help for control panel utilization/access.
Operating principle of the control panel
C
110
Access
to level 4 - Password
C
130
Setting
and
maintenance
part
presentation
of the control panel
Access
to
parameters
and
parameters
active depending on
authorization
C
120
Summary
of screens
Complete displays list with
access
way and
associated
sheet
C
140
Purpose
of the
external calibration
In
some instances,
it may be
convenient
to display a helium leak
value so that
matches
a desired target value (typically the value of an
external calibrated leak
connected
to the installation to
test
or at the
inlet port of the leak detector):
In hard
vacuum test
mode
When the
measurement
range is very different from the value of the
internal calibrated leak.
In sniffing
test
mode
When a specific calibration is required in the sniffing
test
mode to
certify that the
measurement
is valided and accurated.
In hard
vacuum
or sniffing
test
mode
When the leak
detector
is
connected
to an installation having its own
pumping
system
in operation and a small amount of the leak goes
into the leak detector.
The
external calibration allows to get a direct
readout of the
current
leak value.
When the helium signal
needs
to be displayed in a different unit for
convenience
or to calibrate the leak
detector
if the internal calibrated
leak is temporarily unavailable (manual auto-calibration selected).
The
external calibration is provided for the operator to easily obtain
a direct readout of the target value (or
current
external leak)
thanks
to
a correction factor automatically calculated and applied to the
digital
display of the leak detector.
Note : For the ASI 22, the