E
740
Greasing molecular and
turbomolecular pumps
GB 01240
-
Edition 04
- Juin
08
After relubricating, the P0326 board could not allow you to
success
the first or
second
leak
detector
starting. Why ? When you use
a grease syringe, there is a grease
accumulation
on one point of
the ball bearing, and so a small hard point, which will disappear
progressively
during the grease repartition.
What should you do ?
Just
after relubrication,
set
the pump going with
your hand few
minutes
before to replace it (if it was removed from the
leak
detector)
or to take off the flexible (if it was not removed from the
leak
detector)
: in this case, try to start the pump before removing the
flexible : it is not a destroying
test.
You can also
use
a turbo converter
(more powerful for this first critical half turn). After you should
also
keep the leak
detector
running 2 or 3
hours
before stopping
it.
Adjust
the
correspondance
between
the
dif
ferent
test modes
The
internal
parameters
allow a
correspondance
of the different test
modes
depend on the pump.
Thats
why after every
intervention
of the pump it is
indispensable
to
for new adjustment
E
430.
4
/5