"Sniffer
probe
clogged"
message
A “Sniffer probe clogged”
message
could appear on the control panel
LCD
display or be announced by the digital voice: the leak detector
compares
the helium signal read on the
detector
to the sniffer probe
clogged reject threshold.
10
-7
mbar
.l/s
Sniffer
probe clogged message No message
10
-5
mbar
.l/s
Sniffer
probe clogged
reject
threshold value
Value of the
helium
signal
read on the detector
During the needle
adjustment,
this
message
could appear without the
snifffer probe
necessarily
being clogged: this is why the needle length
is so important.
For more details, please
consult
the
users
manual delivered with your
leak detector.
Advice:
Block
the sniffer probe end from time to time with a finger to check
that the helium signal goes down. If not, the sniffer probe may be
clogged.
Adaptor for
calibrated
leak
Reference
part number
F
1000
Special adaptors for calibrated
leaks have been designed to
ensure
a good
connection
and repetitive
and reliable calibration with
a
sniffer probe.
Adaptor
Calibrated
leak
Sniffer
probe
With the adaptor for calibrated leaks use:
calibrated leak
value
Value read on the leak
detector
=
{
+
value of the helium in the
air
8
/8