Advanced internal calibration of the leak
detector
GB 00970 -
Edition
05 -
June
08
Checking
function
Internal calibration activated (on), checking function
performs
a
calibration checking depending on
set
parameters.
The
calibration checking is performed with the internal calibrated leak
of the leak
detector
(position parameter = internal).
C
305
On
request
calibration
checking
At any time, the operator could perform a calibration checking with
the internal calibrated leak.
STDBY
+
Time
between
both < 5
s.
The
leak
detector compares
the signal delivered by the internal
calibrated leak to the internal calibrated leak
parameters
set:
if the difference is lower than 15 %, the calibration of the leak
detector
is
OK.
if the difference is higher than 15 %, there is a warning message
which
requests
to perform a
complete
calibration of the leak detector.
C
301
Parameters
setting
Adjustment
C
110
Automatic
Operator
Adjustment
Adjustment
Cycles
counter
Hours counter
If checking function is automatic, operator should adjust
cycles
and
hours counters
which will trigger the automatic calibration checking.
Note:
The
first of the 2
counters
reached will trigger the automatic
checking.
Tickets If a printer is
connected
to the leak detector, a ticket is automatically
printed at the end of the calibration: example