400
Basic maintenance
of the analyzer cell
GB 00046 -
Edition
07 -
June
08
The frequencies
of preventive
maintenance tasks
are
listed in section D
100.
Components
F
500
Safety
instructions
B
100
Special
precautions
CAUTION
Disconnect
the
detector
from the main powe
r
.
The
analyzer cell is very
sensitive
to any form of
contamination, particularly to dust and
electrostatic
discharge.
When
assembling,
to avoid gettering due to dust or finger prints, you
are advised to work:
in a clean room,
on lent free paper
,
with unpowered vinyl gloves (clean room gloves),
to blow off each part with filtered dry
air
,
to block all the openings in the
vacuum
lines and the VHS
preamplifier
.
The
operator
must
take all
necessary measures
to avoid transferring
electrostatic charges
during the operation.