Cleaning Tips
To access the OPM Port
Unscrew the adapter cap from the adapter cap mount.
Cleaning the Exposed Ferrule or the OPM port
Use lint-free optical cleaning wipes such as AFL FiberWipes and optical quality cleaning uid such
as AFL FCC2 connector cleaning uid.
Note: If using isopropyl alcohol (IPA), be sure to use 99% pure IPA that has not been
contaminated.
Dampen a portion of the wipe with the cleaning uid.
Gently wipe the exposed ferrule (OPM port) starting with the wet section of the wipe and
pulling it to the dry section.
Note: Starting with the wet cleaning and nishing in the dry improves cleaning action,
reduces static buildup, and nishes with the end-face dry.
Cleaning the adapters
Method 1:
Insert a Cletop adapter cleaning stick into the sleeve of the adapter and rotate 10 times.
Remove.
After cleaning the adapter, replace the adapter over the ferrule; centering it onto the alignment
pin.
Tighten the adapter base.
Method 2:
Use a can of ltered compressed air (held vertically), blow out any contaminates from the
adapter.
After cleaning the adapter, replace the adapter over the ferrule; centering it onto the alignment
pin.
Tighten the adapter base.