Getting on and off the table
WARNING:
If the tabletop is moved to the left or to the right, it is prohibited
to let the patient get on or off the tabletop at the protruding end
of the tabletop.
1 1
22 3
1.
Protruding end of the tabletop
2.
Prohibited to get on or off the table
3.
Allowed to get on or off the table
WARNING:
With a patient weighing more than 265 kg (up to max. 320 kg),
the radiographic table must be in the lowest position and not
moved while it is under load.
For patient weights higher than 265 kg (up to max. 320 kg) it is necessary to
place the table at its minimum height, to position the mobile area based on
the area to be examined and, only later, to allow the patient on the tabletop.