EasyManua.ls Logo

AGFA DR 800 - Patient Size

AGFA DR 800
322 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Size
The size of the patient is classified in five categories: Extra Small, Small,
Medium, Large and Extra Large.
Touch the UP or DOWN arrows to select the desired patient size.
Table 33: kV variation over patient size
Patient size kV
Extra Small normal kV * 0.9
Small normal kV * 0.95
Medium normal kV
Large normal kV * 1.05
Extra Large normal kV * 1.1
Patient Size using two point mode
When using two point mode, the patient size setting affects the mAs value.
Table 34: mAs variation over patient size
Patient size mAs
Extra Small normal mAs * 0.25
Small normal mAs * 0.5
Medium normal mAs
Large normal mAs * 2
216 | DR 800 | Software Console