Anatomical stitching is not optimal
Table 48: Problem: Anatomical stitching is not optimal
Details The anatomic information in the partial images cannot
be automatically fully aligned. If the result of automatic
stitching based on the alignment of the anatomical in-
formation in the image is suspicious because of possible
movement of the patient during the examination, this
icon is displayed in the Stitching pane:
Cause
The patient has changed position during the examina-
tion.
Brief Solution Manually adjust the DR Full Leg Full Spine image.
If the partial images cannot be manually adjusted, click
Cancel in the Stitching pane. No DR Full Leg Full Spine
image is available.
DR 800 | Problem solving | 267
0392C EN 20210309 1049