Full Leg Full Spine examination is interrupted
Table 50: Problem: Examination is interrupted
Details The DR Full Leg Full Spine examination is aborted before it has
completely finished.
Cause The exposure button is released by the user before the complete
examination has finished.
Brief Sol-
ution
If you release the exposure button by accident, you can press it
again to continue the examination.
If the exposure button remains released for more than 2 seconds,
the examination is aborted. You can use the available partial im-
ages to create a DR Full Leg Full Spine image.
DR 800 | Problem solving | 271