One point, two point and three point working modes
You can select following radiographic working modes according to the
parameters to be controlled and the degree of automation:
Table 28: Radiographic working modes
mAs mA,ms
One point mode, by selecting kV. The
exposure is controlled by AEC.
mAs mA,ms
Two point mode, by selecting kV and
mAs. AEC is disabled.
mAs mA,ms
Three point mode, by selecting kV,
mA and exposure time independent-
ly. AEC is disabled.
To switch to one point mode, activate one or more AEC fields.
Depending on the radiographic working mode, some of the generator controls
will be disabled.
208 | DR 800 | Software Console