Density
These buttons are used to adjust the AEC cut-off dose (and patient entrance
dose accordingly).
Density can be increased and decreased in a range of -4 to +4, representing a
change of -50% to +100% in dose. When disabled, the density range number
appears in black.
Table 32: Dose variation compared to reference dose
(D)
Dose
-4 0.50
-3 0.63
-2 0.75
-1 0.88
0 1 (reference dose)
+1 1.25
+2 1.50
+3 1.75
+4 2.00
DR 800 | Software Console | 215
0392C EN 20210309 1049