Remarks for HF-emission and immunity
(radiographic wall stand)
Annex A
Guidance and manufacturer’s declaration – Electromagnetic emissions
The system is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should assure that it is used in
such an environment.
Transmission Measure-
ments
Agreement Electromagnetic Environment
Guidelines
RF emissions in accord-
ance with CISPR 11
Group 1 The system uses RF energy only
for its internal function. There-
fore, its RF emissions are very
low and are not likely to cause
any interference in nearby elec-
tronic equipment.
RF emissions in accord-
ance with CISPR 11
Class A The system is suitable for use in
all establishments other than do-
mestic and those directly con-
nected to the public low-voltage
power supply network that sup-
plies buildings used for domestic
purposes.
Harmonic emissions in
accordance with IEC
61000-3-2
Not applicable
Voltage fluctuations /
flicker emissions in ac-
cordance with
IEC 61000-3-3
Not applicable
Annex B
Guidance and manufacturer’s declaration – Electromagnetic immunity
The system is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should assure that it is used in
such an environment.
Immunity Test IEC 60601-1-2
Test level
Compliance
level
Electromagnetic
Environment
Discharge of static
electricity in ac-
cordance with IEC
61000-4-2
6 kV contact
8 kV air
IEC 60601-1-2
Test level
Floors should con-
sist of wood, con-
crete or ceramic
tiles. The relative
humidity must be
DR 800 | Product Information | 315
0392C EN 20210309 1049