at least 30%, if the
floor is made of
synthetic material.
Fast transient elec-
trical disturbance
variables / bursts
in accordance with
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/
ouput lines > 3
m
IEC 60601-1-2
Test level
The quality of the
voltage supplied
should correspond
to a typical com-
mercial or clinical
environment.
Impulse voltages
(surges) in accord-
ance with IEC
61000-4-5
1 kV differential
mode
2 kV common
mode
IEC 60601-1-2
Test level
The quality of the
voltage supplied
should correspond
to that of a typical
commercial or
clinical environ-
ment.
Voltage dips, short
interruptions and
voltage variations
on power supply
iput lines in ac-
cordance with IEC
61000-4-11
0% U
n
for 0.5 cy-
cle
40% U
n
for 5 cy-
cles
70% U
n
for 25 cy-
cles
0% U
n
for 5 s
IEC 60601-1-2
Test level
Mains power qual-
ity should be that
of a typical com-
mercial or hospital
environment. If
the user of the sys-
tem requires con-
tinued operation
during power
mains interrup-
tions, it is recom-
mended that the
system be pow-
ered from an unin-
terruptible power
supply or a bat-
tery.
Magnetic field at
the supply fre-
quency (50/60
Hz) in accordance
with IEC
61000-4-8
3 A/m
IEC 60601-1-2
Test level
Magnetic field at
the network fre-
quency should cor-
respond to the typ-
ical values as they
are in a commer-
cial and clinical
environment.
REMARK : U
n
is the alternating current in the network before the appli-
cation of the test level.
316 | DR 800 | Product Information
0392C EN 20210309 1049