AGFA DR 800 - Immunity to RF Wireless Communication Equipment

AGFA DR 800
322 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Voltage dips,
short interrup-
tions and voltage
variations on
power supply iput
lines in accord-
ance with IEC
61000-4-11
0% U
n
for 0.5 cy-
cle
0% U
n
for 1 cycle
70% U
n
for 25
cycles
0% U
n
for 5 s
IEC 60601-1-2
Test level
Mains power quality
should be that of a
typical commercial
or hospital environ-
ment. If the user of
the system requires
continued operation
during power mains
interruptions, it is
recommended that
the system be pow-
ered from an unin-
terruptible power
supply or a battery.
Magnetic field at
the supply fre-
quency (50/60
Hz) in accordance
with IEC
61000-4-8
30 A/m
IEC 60601-1-2
Test level
Magnetic field at the
network frequency
should correspond
to the typical values
as they are in a com-
mercial and clinical
environment.
REMARK : U
n
is the alternating current in the network before the appli-
cation of the test level.
Immunity to RF wireless communication equipment
ISM Band
(MHz)
Service
Distance
(m)
Immunity
test level
(V/m)
300-390 TETRA 400 0.3 27
430-470 GMRS 460; FRS 460 0.3 28
704-787 LTE Band 13, 17 0.3 9
800-960 GSM 800/900; TETRA 800, IDEN
820; COMA 850; LTE Band 5
0.3 28
1700-1990 GSM 1800; COMA 1900; GSM 1900;
DECT; LTE Band 1, 3, 4, 25; UMTS
0.3 28
2400-2570 Bluetooth; WLAN; 802.11 b/g/n;
RFID 2450; LTE Band 7
0.3 28
5100-5800 WLAN 802.11 a/n 0.3 9
314 | DR 800 | Product Information
0392C EN 20210309 1049