Fixed DR Detector Technical Data
Manufacturer
Manufacturer DR Detector THALES AVS FRANCE SAS
460 Rue du Pommarin – BP122
38430 MOIRANS
France
Supported model
Pixium RF4343 FL
Warming-up time
270 minutes
Reliability
Estimated product life (if regularly serviced
and maintained according to Agfa instruc-
tions)
minimum 10 years or 100 Gy
Pixel Matrix
Pixel size 148 µm
Pixel matrix 2840 x 2874
Active area size 43 x 43 cm
Topics:
• Fixed DR Detector environmental conditions
• Dynamic imaging modes
Fixed DR Detector environmental conditions
Pixium RF4343 FL
Environmental Conditions (during normal operation)
Temperature
Humidity
Atmospheric pressure
Refer to environmental conditions of the X-ray
system
292 | DR 800 | Product Information
0392C EN 20210309 1049