Fluo timer
The fluo timer displays the total duration up till now of all fluoroscopy
sequences in this examination.
Figure 153: Fluo timer
To prevent excessive fluoroscopy time, an alarm time and a maximum
duration time can be configured.
When the alarm time is reached, a sound signal is heard and the fluo timer is
blinking orange.
When the maximum duration time is reached, the fluoroscopy sequence is
stopped.
There can be a difference of a few seconds between the value of the fluo timer
in the software console and the value of the total duration of all dynamic
images in the NX software.
202 | DR 800 | Software Console