Maintenance of the ceiling suspension and the radiographic
wall stand
WARNING:
The steel cable and winding drums must be checked every year.
The steel cable must be replaced following the indication on the
label that is placed on the equipment and in the latest every 5
years for the ceiling suspension and every 10 years for the wall
stand.
WARNING:
The manufacturer declines every responsibility for any kind of
negligence in following the instructions. The lack of check and
controls may cause the system to fall and represents a serious
threat to life for both patient and operator.
WARNING:
A broken spring causes a deficit of several kilograms of force in
the telescope balancing system. The braking system will
compensate the force and keep the movement stable. The ceiling
suspension must not be used until the broken spring is replaced.
Table 6: Maintenance
Periodic maintenance
All steel cables, winding drums, pulleys and
safety devices of the ceiling suspension and
the radiographic wall stand shall be checked
Every 12 months
Condition of the tension springs of the ceil-
ing suspension.
Every 12 months
Replacement of steel cables of the ceiling
suspension
Every 5 years, indicated on
the label that is placed on the
equipment
Replacement of steel cables of the wall stand Every 10 years, indicated on
the label that is placed on the
equipment
Replacement of tension springs of the ceiling
suspension
As needed