Tests of Re-
sistance to
Disruption
IEC
60601-1-2:2014
Test Level
Level of
Agree-
ment
Electromagnetic Environ-
ment
Conducted
RF
IEC
61000-4-6
3 V
150 kHz to 80 MHz
3 V Portable and mobile RF com-
munications equipment
should be used no closer to
any part of the device, includ-
ing cables, than the recom-
mended separation distance
calculated from the equation
applicable to the frequency of
the transmitter.
Recommended separation
distance:
d = 1.2
150 kHz to 80
MHz
d = 1.2 80 MHz to 800
MHz
d = 2.3
800 MHz to 2.5
GHz
where ’P’ is the maximum out-
put power rating of the trans-
mitter in watts (W) according
to the transmitter manufac-
turer and ’d’ is the recommen-
ded separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, should be less than
the compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Radiated RF
IEC
61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Spot fre-
quencies
DR 800 | Product Information | 317
0392C EN 20210309 1049