Using the Detectors
Configuring the Analog Channel Output
Wavelength Detector and Refractive Index Detector have one) at the back side of the
module.
To configure the analog outputs:
1 Press the Settings button [F1] in the Analysis screen.
2 Choose the detector and press enter.
3 Press the Analog button [F3].
4 Either select the signal you want to monitor from the signal source drop down list.
Or, if you have a variable wavelength detector (fluorescence light detector), choose
the signal(s) from the list(s) to be routed to the analog output(s).
5 Enter the values for Zero Offset (sets a baseline offset in order to recognize negative
drifts) and Attenuation (sets an absorbance range) for each of the two connectors.
6 Press the Done button [F6].