Running Multiple-Vial Analyses
Single-Level Calibration Sequences
7 Select the Done button [F6] to enter sequence line 3 settings and complete the
sequence.
8 Select the Start button [F8] in the Sequence screen.
9 Select the OK button to start the sequence.
Single-Level Calibration Sequences
The following procedure describes how to set up a calibration sequence for an
analysis which uses single-level calibration.
There is one calibration standard (C) and 9 samples (S).
The analysis requires that:
each sample is analyzed in duplicate,
the calibration standard is analyzed once before the samples and re-analyzed once
after every 2 samples,
CSSCSSCSSCSSCSC
the calibration standard is located in position 90 and the 9 sample vials are in
positions 1 to 9 of the autosampler tray, and
the method called METH1 is used for the samples and standards.
The example assumes that the sequence table is empty. If the sequence table is not
empty, use the All Lines item, accessed by pressing the Delete button [F6].
1 Select the Sequence button [F4] in the Analysis screen.
2 For sequence line 1, select the Insert button [F7] and enter:
Vial Range 1 to 9
#Inj. 2
Inj. Volume Default
Method METH1

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals