Running Multiple-Vial Analyses
Analyzing Multiple Vials Using Different Methods
The example assumes that the sequence table is empty. If the sequence table is not
empty, use the All Lines item, accessed by pressing the Delete button [F6].
1 Select the Sequence [F4] in the Analysis screen.
2 For sequence line 1, select the Insert button [F7] and enter:
Vial Range 1 to 20
#Inj. 1 1
Inj. Vol. Default
Method Meth1
3 Select the Enter button to accept entries.
4 For sequence line 2, enter:
Vial Range 21 to 40
#Inj. 2
Inj. Volume Default
Method METH2
5 Select the Enter button to accept entries.
6 For sequence line 3, enter:
Vial Range 41 to 50
#Inj. 3
Inj. Volume Default
Method METH3
Wait 30 minutes