72
Using the Degasser
Starting the Degasser
Starting the Degasser
1 Press the line-power switch on the front panel.
2 Turn on your pump by selecting the On/Off button [F8] in the Analysis screen.
3 Observe the degasser status lamp on the front panel.
The status lamp is off when there is sufficient vacuum in the degasser.
During operation, the status lamp may be yellow for several seconds. This indicates
there is insufficient vacuum in the online degasser and the vacuum pump is on to
create sufficient vacuum.
Removing Gas Bubbles
If you see gas bubbles in the solvent tubing or inside the solvent filter:
1 Disconnect the solvent tube of the first solvent channel from your pump.
2 Connect the syringe adapter onto the syringe from degasser accessory kit.
3 Pull the syringe plunger to draw solvent through the degasser and tubing. Continue
to draw solvent through tubing until no gas bubbles are visible.
4 Disconnect the syringe adapter from the solvent tube.
5 Connect the solvent tube to your pump.
Repeat step 1 through step 5 for other solvent channels.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals