121
Running Multiple-Vial Analyses
Multiple-Level Calibration Sequences
3 Select the Calibration button [F1] to display the Calibration Settings screen for
sequence line 1 and choose Edit.
4 Enter the following information:
Calibration vial
range 90 to 90
Number of
injections 1
Calibration
method METH1
Recalibrate every 2 vials multi
Before On
After On
5 Press the Done button [F6] until the Sequence screen appears.
6 Select the Start button [F8] in the Sequence screen.
7 Press the Enter key to start the sequence.
Multiple-Level Calibration Sequences
The following sections describe how to set up calibration sequences for analyses
which use multiple-level calibration.
Recalibrating With the Same Group of
Standards
There are three calibration standards of different concentrations (C1, C2, C3) and 15
samples (S). The standards and samples are analyzed using the same method.
The analysis requires that:

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals