121
Running Multiple-Vial Analyses
Multiple-Level Calibration Sequences
3 Select the Calibration button [F1] to display the Calibration Settings screen for 
sequence line 1 and choose Edit.
4 Enter the following information:
Calibration vial 
range  90 to 90
Number of 
injections  1
Calibration 
method  METH1
Recalibrate every  2 vials multi
Before   On
After   On
5 Press the Done button [F6] until the Sequence screen appears.
6 Select the Start button [F8] in the Sequence screen.
7 Press the Enter key to start the sequence.
Multiple-Level Calibration Sequences
The following sections describe how to set up calibration sequences for analyses 
which use multiple-level calibration.
Recalibrating With the Same Group of 
Standards
There are three calibration standards of different concentrations (C1, C2, C3) and 15 
samples (S). The standards and samples are analyzed using the same method. 
The analysis requires that: