Using the Degasser
Changing Solvents
Changing Solvents
If you are changing to a solvent that is immiscible with the solvent currently in the
tubing:
1 Replace the current solvent:
with iso-propanol, if current solvent is organic, or
with water, if current solvent is an organic buffer or salt.
2 Flush the online degasser and all tubing thoroughly.
3 Replace the iso-propanol or water with the new solvent.