90
Using the Detectors
Resetting the Baseline
Resetting the Baseline
You can reset the baseline using the Balance button [F1] (Zero button [F1] when
using the Refractive Index Detector) in the Detector Control screen. Please note that
this procedure is not available for the Fluorescence Light Detector:
1 Select the Control button [F1] in the System screen.
2 Select the Detector from the menu.
3 Press the Enter key.
4 Select the Balance button [F1].
You can achieve the same using the m (menu) key in the Detector Settings window
available from the Analysis screen.
Configuring the Detector
1 Select the Configure button [F2] in the System screen.
2 Select the Detector from the list.
3 Press the Enter key.
You can now configure the detector analog output(s), lamp-on at power-on and the
interfaces.