EasyManua.ls Logo

Agilent Technologies 1100 Series - Run Recovery

Agilent Technologies 1100 Series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
76
Using the Autosampler
Configuring the Autosampler
Configuring the Autosampler
1 Select the Configure button [F2] in the System screen.
2 Select Autosampler from the menu.
3 Press the Enter key.
You can configure the syringe volume, seat capillary volume, action on missing vial,
trays, interfaces etc.
Configuring Interfaces
1 Select the Configure button [F2] in the System screen.
2 Select Autosampler from the menu.
3 Press the Enter key.
4 Select the Interfaces button [F1].
5 Select the interfaces you want to operate (see Connecting External Devices on
page 51).
6 Select the Done button [F6].
Entering Settings
All autosampler settings can be entered in the Autosampler Settings screen. You can
access this screen by selecting the Settings button [F1] in the Analysis screen. From
here you can access the timetable function (to set up a timetable for the Autosampler
only), the runtimes screen and the injector program screen. Commonly-used settings
such as injection volume and vial number can be set directly in the Analysis screen.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals