122 Agilent 1260 Infinity DAD and MWD User Manual
6 How to optimize the Detector
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Tab le 29 Peak Width — Response Time — Data Rate
Peak Width
[minutes]
Response Time
[seconds]
Data Rate [Hz] Detector
<0.0025 0.025 80 G1315C/G1365C only
>0.0025 0.05 80 G1315C/G1365C only
>0.005 0.1 40 G1315C/G1365C only
>0.01 0.2 20 G1315C/D and G1365C/D
>0.03 0.5 10 G1315C/D and G1365C/D
>0.05 1.0 5 G1315C/D and G1365C/D
>0.10 2.0 2.5 G1315C/D and G1365C/D
>0.20 4.0 1.25 G1315C/D and G1365C/D
>0.40 8.0 0.62 G1315C/D and G1365C/D
>0.85 16.0 0.31 G1315C/D and G1365C/D