EasyManua.ls Logo

Agilent Technologies 1260 - Page 125

Agilent Technologies 1260
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1260 Infinity DAD and MWD User Manual 125
How to optimize the Detector
6
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Because the detector averages absorbance values that are calculated for each
wavelength, using a wide bandwidth does not negatively impact linearity.
The use of a reference wavelength is highly recommended to further reduce
baseline drift and wander induced by room temperature fluctuations or
refractive index changes during a gradient.
An example of the reduction of baseline drifts is shown in Figure 55 on
page 125 for PTH-amino acids. Without a reference wavelength, the
chromatogram drifts downwards due to refractive index changes induced by
the gradient. This is almost completely eliminated by using a reference
wavelength. With this technique, PTH-amino acids can be quantified in the low
picomole range even in a gradient analysis.
Figure 55 Gradient Analysis of PTH-Amino Acids (1 pmol each), with and without
Reference
I^bZb^c
<gVY#/%#%'b@=ED$68C[gdb&'68Cid)*68C^c&'b^c
&ebdaZVX]
GZ[ZgZcXZ(-%cb
LVkZaZc\i]
'+,cb
LVkZaZc\i]
'+,cb
CdgZ[ZgZcXZ
EI="6HC
EI="6G<
EI="6A6
EI="E=:
EI="EGD
')

Table of Contents

Other manuals for Agilent Technologies 1260

Related product manuals