Agilent 1260 Infinity DAD and MWD User Manual 219
Maintenance
10
Nano Flow Cell - Replacing or Cleaning
1
Disconnect the capillaries from the capillary holder and
remove the flow cell.
2 Unscrew the cell body from the holder.
3 Unscrew the capillaries from the flow cell. DO NOT use
the adapter at this time!
4 Using for example a toothpick, press on the plastic part
and slide the quartz body out of the cell housing.