© Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-110
Chapter 10: Site Preparation – Power Requirements for Older Systems: Tester Power Requirements (Series I/II)
Table 10-18 Identifying PDU / PDK type
307X Series I System:
Shipped before 1/1/92: 44964A/B PDU See Figure 10-69 on page 10-144
Shipped 1/1/92 to 12/1/93: E1099A PDU See Figure 10-70 on page 10-145
Shipped after 12/1/93: E1131A PDU( See Figure 10-75 on page 10-155
317X Series I System:
All ship dates: E1170-80003 PDK See Figure 10-72 on page 10-149
3X7X Series II System:
Shipped before 7/7/95: E1131A PDU See Figure 10-75 on page 10-155
Shipped 7/7/95 - 11/21/96: E1135A PDU See Figure 10-76 on page 10-159 -
Figure 10-80 on page 10-163
Shipped 11/21/96 - 9/4/98: E1135B PDU See Figure 10-76 on page 10-159 -
Figure 10-80 on page 10-163
3X7X Series 3 System:
Shipped before 9/4/98: E1135B PDU See Figure 10-76 on page 10-159 -
Figure 10-80 on page 10-163
Shipped after 9/4/98: E1135C PDU See Chapter 6, Site Preparation –
Power Requirements
POTS Bay:
Shipped before 7/7/95: has an additional E1131A PDU See Figure 10-75 on page 10-155